4.0 Article

Measuring in vitro ATPase Activity with High Sensitivity Using Radiolabeled ATP

期刊

BIO-PROTOCOL
卷 13, 期 10, 页码 -

出版社

BIO-PROTOCOL
DOI: 10.21769/BioProtoc.4676

关键词

ABC transporter; ATPase; ATP hydrolysis; Radioactive assay; P-32-ATP; P-type ATPase

类别

向作者/读者索取更多资源

ATPase assays are commonly used for characterizing purified ATPases. This radioactive [gamma-P-32]-ATP-based approach described here can be highly sensitive and is suitable for proteins with low ATPase activity or low purification yields. It can be used for various applications including substrate identification, determining the effect of mutations on ATPase activity, and testing ATPase inhibitors.
ATPase assays are a common tool for the characterization of purified ATPases. Here, we describe a radioactive [gamma-P-32]-ATP-based approach, utilizing complex formation with molybdate for phase separation of the free phosphate from non-hydrolyzed, intact ATP. The high sensitivity of this assay, compared to common assays such as the Malachite green or NADH-coupled assay, enables the examination of proteins with low ATPase activity or low purification yields. This assay can be used on purified proteins for several applications including the identification of substrates, determination of the effect of mutations on ATPase activity, and testing specific ATPase inhibitors. Furthermore, the protocol outlined here can be adapted to measure the activity of reconstituted ATPases.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.0
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据