3.8 Article

Comparative study of quality of leukoreduced packed red blood cell units as assessed by nageotte hemocytometry and flow cytometry

期刊

ASIAN JOURNAL OF TRANSFUSION SCIENCE
卷 17, 期 1, 页码 63-68

出版社

WOLTERS KLUWER MEDKNOW PUBLICATIONS
DOI: 10.4103/ajts.AJTS_101_21

关键词

Alloimmunization; flow cytometry; leukodepletion; nageotte hemocytometer; quality control; residual white blood cell count

向作者/读者索取更多资源

This study compared the use of Nageotte hemocytometer and flow cytometry (FC) for quality control of LD red blood cell units. The results showed that FC provides a more precise and accurate objective tool compared to Nageotte hemocytometer, which is a reliable alternative in resource-constrained settings.
PURPOSE: Assessment of residual white blood cell (rWBC) count is vital to ascertain the quality of leukodepleted (LD) blood components. Automated cell analyzers lack the sensitivity for the assessment of very few leukocytes as found in LD blood components. Flow Cytometry (FC) based methods and Nageotte hemocytometer are the most commonly used techniques for this purpose. The objective of this study was to compare the use of Nageotte hemocytometer and FC for quality control of LD red blood cell units. MATERIALS AND METHODS: A prospective, observational study was conducted in the Department of Immunohematology and Blood Transfusion of a tertiary care center from September 2018 to September 2020. About 303 LD-packed red blood cell units were tested by FC and Nageotte hemocytometer for rWBCs. RESULTS: The number of rWBC (mean) detected by flow cytometer and Nageotte's hemocytometer was 1.06 +/- 0.43 white blood cell (WBC)/mu L and 0.67 +/- 0.39 WBC/mu L, respectively. Coefficient of variation was 58.37% by Nageotte hemocytometer method and 40.46% by FC. Linear regression analysis did not show any correlation (R-2 = 0.098, P = 0.001) whereas Pearson's correlation coefficient showed a weak relation (r = 0.31) between the two methods. CONCLUSION: Flow cytometric technique provides a more precise and accurate objective tool compared to Nageotte hemocytometer which is labor intensive, time consuming, and prone to errors arising out of subjectivity along with reported underestimation bias. In the absence of adequate infrastructure, resources, and trained workforce, Nageotte hemocytometer method is a reliable alternative. Nageotte's chamber could be best used in the resource-constrained setup as it offers a relatively inexpensive, simple, and viable means to enumerate rWBCs.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

3.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据