OBJECTIVE To evaluate safety alerts and recalls, publication of key trial outcomes, and subsequent US approval of high profile medical devices introduced in the European Union. DESIGN Cohort study. SETTING Novel cardiovascular, orthopedic, and neurologic devices approved in the EU through Conformite Europeenne marking between 2005 and 2010. DATA SOURCES Public and commercial databases searched up to January 2016 for press releases and announcements of approvals; public Food and Drug Administration and European regulatory authority databases for US approvals and safety alerts and recalls; and Medline, Embase, and Web of Science for peer reviewed publications. MAIN OUTCOME MEASURES We categorized the novelty of the devices in the study sample as a major innovation or an other change, and extracted descriptive data about the devices and information on any safety alerts and withdrawals. Linear regression models examined factors associated with differential EU and US approvals. Cox proportional hazards regression models were used to evaluate factors associated with safety alerts and recalls and the publication of trial outcomes for devices categorized as major innovations. Models controlled for time, therapeutic category, regulatory pathway, size of sponsoring company, and indicator variables for devices approved first in the EU and devices approved only in the EU. RESULTS 67% (206/309) of devices identified were approved in both the US and the EU, of which 63% (129/206) were approved first in the EU. The unadjusted rate of safety alerts and recalls for devices approved first in the EU was 27% (62/232) compared with 14% (11/77) for devices approved first in the US. The adjusted hazard ratio for safety alerts and recalls was 2.9 (95% confidence interval 1.4 to 6.2) for devices approved first in the EU. The results of pivotal trials were published for 49% (37/75) of devices categorized as major innovations, with an overall publication rate of 37% five years after approval. CONCLUSIONS Devices approved first in the EU are associated with an increased risk of post-marketing safety alerts and recalls. Poor trial publication rates mean that patients and clinicians need greater regulatory transparency to make informed decisions about treatment.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据