Clinical Validation of 5 Direct-to-Consumer Wearable Smart Devices to Detect Atrial Fibrillation BASEL Wearable Study

期刊

JACC-CLINICAL ELECTROPHYSIOLOGY
卷 9, 期 2, 页码 232-242

出版社

ELSEVIER
DOI: 10.1016/j.jacep.2022.09.011

关键词

atrial fibrillation; digital health; intelligent ECG; smartwatch

向作者/读者索取更多资源

This study assessed the accuracy of 5 smart devices in detecting atrial fibrillation (AF) compared to physician-interpreted 12-lead electrocardiogram. The sensitivity and specificity varied among the devices, and manual review was required for about one-fourth of the cases.
BACKGROUND Multiple smart devices capable to detect atrial fibrillation (AF) are presently available. Sensitivity and specificity for the detection of AF may differ between available smart devices, and this has not yet been adequately investigated. OBJECTIVES The aim was to assess the accuracy of 5 smart devices in identifying AF compared with a physicianinterpreted 12-lead electrocardiogram as the reference standard in a real-world cohort of patients. METHODS We consecutively enrolled patients presenting to a cardiology service at a tertiary referral center in a prospective, diagnostic study. RESULTS We prospectively analyzed 201 patients (31% women, median age 66.7 years). AF was present in 62 (31%) patients. Sensitivity and specificity for the detection of AF were comparable between devices: 85% and 75% for the Apple Watch 6, 85% and 75% for the Samsung Galaxy Watch 3, 58% and 75% for the Withings Scanwatch, 66% and 79% for the Fitbit Sense, and 79% and 69% for the AliveCor KardiaMobile, respectively. The rate of inconclusive tracings (the algorithm was unable to determine the heart rhythm) was 18%, 17%, 24%, 21%, and 26% for the Apple Watch 6, Samsung Galaxy Watch 3, Withings Scan Watch, Fitbit Sense, and AliveCor KardiaMobile (P < 0.01 for pairwise comparison), respectively. By manual review of inconclusive tracings, the rhythm could be determined in 955 (99%) of 969 single-lead electrocardiograms. Regarding patient acceptance, the Apple Watch was ranked first (39% of participants). CONCLUSIONS In this clinical validation of 5 direct-to-consumer smart devices, we found differences in the amount of inconclusive tracings diminishing sensitivity and specificity of the smart devices. In a clinical setting, manual review of tracings is required in about one-fourth of cases. (c) 2023 Published by Elsevier on behalf of the American College of Cardiology Foundation.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据