4.5 Article

A Comparison Between a Semi-Continuous Analyzer and Filter-Based Method for Measuring Anion and Cation Concentrations in PM10 at an Urban Background Site in London

期刊

AEROSOL SCIENCE AND TECHNOLOGY
卷 49, 期 9, 页码 793-801

出版社

TAYLOR & FRANCIS INC
DOI: 10.1080/02786826.2015.1073848

关键词

-

资金

  1. UK Department for Business Innovation and Skills of the National Measurement System's Chemical & Biological Metrology Programme
  2. UK Department for Environment, Food and Rural Affairs of the UK Particle Number and Concentration Air Quality Network

向作者/读者索取更多资源

A comparison between the filter-based, laboratory ion chromatography technique and a semi-continuous analyzer (URG 9000B Ambient Ion Monitor [AIM]) was conducted to evaluate the performance of the AIM in measuring the concentrations of the main airborne ionic species in PM10. The study was carried out in an urban background area of London (UK) in 2013. The two methods showed an overall good correlation (R-2 > 0.83) for nitrate, sulfate, chloride, ammonium, and magnesium and poor correlation was found for sodium, potassium, and calcium (R-2 < 0.50). The AIM gave consistently higher concentrations for sodium and potassium, possibly due to a positive bias within the sampling unit. During high concentration episodes, both the efficiency of the particle extraction and removal of gases by the denuder may be reduced. A HEPA filter test demonstrated that the denuder was removing gaseous components effectively but that there was some potential for contamination. Overall, the AIM was found to be a good instrument for measuring hourly anion and cation concentrations in PM10 in urban sites.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据