4.6 Article

Pencil beam all-optical ultrasound imaging

期刊

BIOMEDICAL OPTICS EXPRESS
卷 7, 期 9, 页码 3696-3704

出版社

OPTICAL SOC AMER
DOI: 10.1364/BOE.7.003696

关键词

-

资金

  1. European Research Council (ERC) [310970 MOPHIM]
  2. Innovative Engineering for Health award by the Wellcome Trust [WT101957]
  3. Engineering and Physical Sciences Research Council (EPSRC) [NS/A000027/1]
  4. European Union project FAMOS (FP7 ICT) [317744]

向作者/读者索取更多资源

A miniature, directional fibre-optic acoustic source is presented that employs geometrical focussing to generate a nearly-collimated acoustic pencil beam. When paired with a fibre-optic acoustic detector, an all-optical ultrasound probe with an outer diameter of 2 : 5 mm is obtained that acquires a pulse-echo image line at each probe position without the need for image reconstruction. B-mode images can be acquired by translating the probe and concatenating the image lines, and artefacts resulting from probe positioning uncertainty are shown to be significantly lower than those observed for conventional synthetic aperture scanning of a non-directional acoustic source. The high image quality obtained for excised vascular tissue suggests that the all-optical ultrasound probe is ideally suited for in vivo, interventional applications.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据