4.6 Article

Fast blood flow monitoring in deep tissues with real-time software correlators

期刊

BIOMEDICAL OPTICS EXPRESS
卷 7, 期 3, 页码 776-797

出版社

OPTICAL SOC AMER
DOI: 10.1364/BOE.7.000776

关键词

-

资金

  1. National Institutes of Health [R01 - NS060653, 8P41 - EB015893]
  2. Application Foundation and Advanced Technology Research Program of Tianjin [14CZDJC33100]
  3. China Scholarship Council [201306250114]
  4. American Heart Association [14POST20460161]

向作者/读者索取更多资源

We introduce, validate and demonstrate a new software correlator for high-speed measurement of blood flow in deep tissues based on diffuse correlation spectroscopy (DCS). The software correlator scheme employs standard PC-based data acquisition boards to measure temporal intensity autocorrelation functions continuously at 50 - 100 Hz, the fastest blood flow measurements reported with DCS to date. The data streams, obtained in vivo for typical source-detector separations of 2.5 cm, easily resolve pulsatile heart-beat fluctuations in blood flow which were previously considered to be noise. We employ the device to separate tissue blood flow from tissue absorption/scattering dynamics and thereby show that the origin of the pulsatile DCS signal is primarily flow, and we monitor cerebral autoregulation dynamics in healthy volunteers more accurately than with traditional instrumentation as a result of increased data acquisition rates. Finally, we characterize measurement signal-to-noise ratio and identify count rate and averaging parameters needed for optimal performance. (C) 2016 Optical Society of America

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据