4.6 Article

Lock-in-photon-counting-based highly-sensitive and large-dynamic imaging system for continuous-wave diffuse optical tomography

期刊

BIOMEDICAL OPTICS EXPRESS
卷 7, 期 2, 页码 499-511

出版社

OPTICAL SOC AMER
DOI: 10.1364/BOE.7.000499

关键词

-

资金

  1. National Natural Science Foundation of China [81271618, 81371602, 61475115, 61475116, 81401453, 61575140, 81571723]
  2. Tianjin Municipal Government of China [13JCZDJC28000, 14JCQNJC14400, 15JCZDJC31800]

向作者/读者索取更多资源

We implemented a novel lock-in photon-counting detection architecture that combines the ultra-high sensitivity of the photon-counting detection and the measurement parallelism of the lock-in technique. Based on this technique, a dual-wavelength simultaneous measurement continuous wave diffuse optical tomography system was developed with a configuration of 16 sources and 16 detectors that works in a tandem serial-to-parallel fashion. Methodology validation and performance assessment of the system were conducted using phantom experiments that demonstrate excellent measurement linearity, moderate-term system stability, robustness to noise and negligible inter-wavelength crosstalk. 2-D imaging experiments further validate high sensitivity of the lock-in photon-counting methodology as well as high reliability of the proposed system. The advanced detection principle can be adapted to achieving a fully parallelized instrumentation for the extended applications. (C) 2016 Optical Society of America

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据