4.3 Article

Multiplexed longitudinal measurement of protein biomarkers in DBS using an automated SISCAPA workflow

期刊

BIOANALYSIS
卷 8, 期 15, 页码 1597-1609

出版社

FUTURE SCI LTD
DOI: 10.4155/bio-2016-0059

关键词

DBS; longitudinal; monoclonal antibodies; MS; protein biomarkers; SISCAPA

向作者/读者索取更多资源

Background: The use of DBS for quantitative protein biomarker measurement has been hindered by issues associated with blood hematocrit variations and lack of detection sensitivity, particularly when multiple biomarkers are measured. Materials & methods: An automated, multiplexed SISCAPA analysis was used to normalize blood volume variations in DBS and quantify proteins of varying abundance in longitudinal specimens. Conclusion: The results showed that after normalizing the spot-to-spot hematocrit variations, peptide surrogates of protein biomarkers could be accurately quantitated in DBS. This allowed the establishment of baselines for a variety of biomarkers in multiple individuals and enabled detection of changes over time, thus offering an effective solution for longitudinal personal monitoring of biomarkers relevant in health and disease.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据