4.2 Article

First Experiences with the New Enterprise2A (R) Stent

期刊

CLINICAL NEURORADIOLOGY
卷 28, 期 2, 页码 201-207

出版社

SPRINGER HEIDELBERG
DOI: 10.1007/s00062-016-0545-9

关键词

Aneurysm; Endovascular; Stent; Enterprise; Coiling

向作者/读者索取更多资源

Stent-assisted coiling of wide-necked intracranial aneurysms has become an established treatment and has significantly benefited from the introduction of compliant, self-expanding devices, such as the Enterprise VRD (EP-VRD). We report our experiences with the successor model, the Enterprise2 (EP2) stent in stent-assisted coiling as well as in the treatment of atherosclerotic stenosis. In 11 consecutive patients 12 EP2 were used to treat 9 intracranial aneurysms and 2 stenoses. All stents could easily be delivered including partial or complete recapturing when necessary. In two cases with sharp angled curves in the carotid siphon there was kinking and flattening of the stent resulting in incomplete wall apposition of the stent. Moreover, when vascular anatomy showed curves with angles > 50A degrees it was regularly observed that the proximal stent markers were asymmetrically arranged along the vessel circumference without influence on the stent apposition. Both findings could be reproduced in a silicone flow model. The EP2 performed well in our small patient cohort; however, above a critical acute angle there may be incomplete wall apposition of the stent.