4.6 Article

Performance criteria for reference measurement procedures and reference materials

期刊

CLINICAL CHEMISTRY AND LABORATORY MEDICINE
卷 53, 期 6, 页码 899-904

出版社

WALTER DE GRUYTER GMBH
DOI: 10.1515/cclm-2015-0104

关键词

bias; calibration; certified reference materials; equivalent measurement results; metrological traceability; uncertainty

向作者/读者索取更多资源

The concept of metrological traceability of measurement results to property values assigned to measurement standards of higher metrological order or to the International System of Units (SI) through sequential calibrations, using reference materials and reference measurement procedures, plays a key role in ensuring that end user measurement procedures perform at an acceptable level in the clinical context. The aim is that measurement results produced over time or by different end users or with different end user measurement procedures for the same measurand will be equivalent within their corresponding uncertainties. These goals can only be reached under certain conditions and if requirements laid down in international standards on calibration concepts, reference measurement procedures and reference materials are fulfilled. Calibration hierarchies have to be implemented correctly and parameters contributing to measurement uncertainty and systematic bias need to be controlled and eliminated, respectively, by technically improving methods and reference materials and intermediate calibrators used for effectively achieving equivalence of measurement results and for meeting analytical performance requirements for in vitro diagnostic devices.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据