4.7 Article

In vivo demonstration of ultrasound power delivery to charge implanted medical devices via acute and survival porcine studies

期刊

ULTRASONICS
卷 64, 期 -, 页码 1-9

出版社

ELSEVIER
DOI: 10.1016/j.ultras.2015.07.012

关键词

Ultrasound powering; Transcutaneous energy transfer; Battery recharging; Implanted batteries; Animal studies

资金

  1. National Institutes of Health - United States National Institute of Bioimaging and Bioengineering [R43EB007421, R44EB007421]

向作者/读者索取更多资源

Animal studies are an important step in proving the utility and safety of an ultrasound based implanted battery recharging system. To this end an Ultrasound Electrical Recharging System (USER (TM)) was developed and tested. Experiments in vitro demonstrated power deliveries at the battery of up to 600mW through 10-15 mm of tissue, 50m W of power available at tissue depths of up to 50 mm, and the feasibility of using transducers bonded to titanium as used in medical implants. Acute in vivo studies in a porcine model were used to test reliability of power delivery, temperature excursions, and cooling techniques. The culminating five-week survival study involved repeated battery charging, a total of 10.5 h of ultrasound exposure of the intervening living tissue, with an average RF input to electrical charging efficiency of 20%. This study was potentially the first long term cumulative living-tissue exposure using transcutaneous ultrasound power transmission to an implanted receiver in situ. Histology of the exposed tissue showed changes attributable primarily due to surgical implantation of the prototype device, and no damage due to the ultrasound exposure. The in vivo results are indicative of the potential safe delivery of ultrasound energy for a defined set of source conditions for charging batteries within implants. (C) 2015 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据