4.7 Article

Fluorescence analyzer based on smartphone camera and wireless for detection of Ochratoxin A

期刊

SENSORS AND ACTUATORS B-CHEMICAL
卷 232, 期 -, 页码 462-468

出版社

ELSEVIER SCIENCE SA
DOI: 10.1016/j.snb.2016.03.140

关键词

Smartphone; Wireless; IP address; Ochratoxin A; Fluorescence analysis and camera

资金

  1. European project COCOtA: Ministere de l'agriculture, de l'agroalimentaire et de la foret [2013e2015]

向作者/读者索取更多资源

This manuscript describes an innovative approach to quantify the concentrations of ochratoxin A (OTA) using the smartphone as fluorescence device. The detection through the smartphone is based on the fluorescence phenomenon. It occurs when the sample is excited with ultraviolet (UV) light, the fluorescence from excited sample passes through a lens to smartphone camera. Finally, it is send by wireless to personal computer. The fluorescence image data from the smartphone camera is analyzed by a personal computer; then, the images are represented in its Red, Green and Blue (RGB) components. It is reported that OTA is naturally fluorescent, hence when the light is emitted through solution with OTA; it provides a blue fluorescence, so the image is captured with the smartphone camera whereas for blank solution, no fluorescence was observed. To process the image, a graphical user interface (GUI) in MATLAB R2011a was made. The best linear range with the developed system was in the range 2-20 mu g/L and the limit of detection (LOD) was 2 mu g/L. This is comparable with commercial equipment in a portable system and employing smartphone as a power source also. (C) 2016 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据