4.6 Article

Development and clinical introduction of automated radiotherapy treatment planning for prostate cancer

期刊

PHYSICS IN MEDICINE AND BIOLOGY
卷 61, 期 24, 页码 8587-8595

出版社

IOP PUBLISHING LTD
DOI: 10.1088/1361-6560/61/24/8587

关键词

automation; treatment planning; prostate cancer; monaco; intensity-modulated radiation therapy

资金

  1. Elekta AB, Stockholm, Sweden

向作者/读者索取更多资源

To develop an automated radiotherapy treatment planning and optimization workflow to efficiently create patient specifically optimized clinical grade treatment plans for prostate cancer and to implement it in clinical practice. A two-phased planning and optimization workflow was developed to automatically generate 77Gy 5-field simultaneously integrated boost intensity modulated radiation therapy (SIB-IMRT) plans for prostate cancer treatment. A retrospective planning study (n = 100) was performed in which automatically and manually generated treatment plans were compared. A clinical pilot (n = 21) was performed to investigate the usability of our method. Operator time for the planning process was reduced to <5 min. The retrospective planning study showed that 98 plans met all clinical constraints. Significant improvements were made in the volume receiving 72Gy (V72Gy) for the bladder and rectum and the mean dose of the bladder and the body. A reduced plan variance was observed. During the clinical pilot 20 automatically generated plans met all constraints and 17 plans were selected for treatment. The automated radiotherapy treatment planning and optimization workflow is capable of efficiently generating patient specifically optimized and improved clinical grade plans. It has now been adopted as the current standard workflow in our clinic to generate treatment plans for prostate cancer.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据