4.6 Article

Bismuth germanate coupled to near ultraviolet silicon photomultipliers for time-of-flight PET

期刊

PHYSICS IN MEDICINE AND BIOLOGY
卷 61, 期 18, 页码 L38-L47

出版社

IOP PUBLISHING LTD
DOI: 10.1088/0031-9155/61/18/L38

关键词

positron emission tomography (PET); time of flight (TOF); Cerenkov (Cherenkov); scintillation; silicon photomultiplier (SiPM); bismuth germanate (BGO)

资金

  1. NIH National Cancer Institute [R35 CA197608]
  2. NIH National Institute of Biomedical Imaging and Bioengineering [R01 EB019439]

向作者/读者索取更多资源

Bismuth germanate (BGO) was a very attractive scintillator in early-generation positron emission tomography (PET) scanners. However, the major disadvantages of BGO are lower light yield and longer rise and decay time compared to currently popular scintillators such as LSO and LYSO. This results in poorer coincidence timing resolution and it has generally been assumed that BGO is not a suitable scintillator for time-of-flight (TOF) PET applications. However, when a 511 keV photon interacts in a scintillator, a number of Cerenkov photons are produced promptly by energetic electrons released by photoelectric or Compton interactions. If these prompt photons can be captured, they could provide a better timing trigger for PET. Since BGO has a high refractive index (increasing the Cerenkov light yield) and excellent optical transparency down to 320 nm (Cerenkov light yield is higher at shorter wavelengths), we hypothesized that the coincidence timing resolution of BGO can be significantly improved by efficient detection of the Cerenkov photons. However, since the number of Cerenkov photons is far less than the number of scintillation photons, and they are more abundant in the UV and blue part of the spectrum, photosensors need to have high UV/blue sensitivity, fast temporal response, and very low noise in order to trigger on the faint Cerenkov signal. In this respect, NUV-HD silicon photomultipliers (SiPMs) (FBK, Trento, Italy) are an excellent fit for our approach. In this study, coincidence events were measured using BGO crystals coupled with NUV-HD SiPMs. The existence and influence of Cerenkov photons on the timing measurements were studied using different configurations to exploit the directionality of the Cerenkov emissions. Coincidence resolving time values (FWHM) of similar to 270 ps from 2 x 3 x 2 mm(3) BGO crystals and similar to 560 ps from 3 x 3 x 20 mm3 BGO crystals were obtained. To our knowledge, these are the best coincidence resolving time values reported for BGO to date. With these values, BGO can be considered as a relevant scintillator for TOF PET scanners, especially if photodetectors with even better near UV/blue response can be developed to further improve the efficiency of Cerenkov light detection.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据