Safety, biocompatibility, and potential functionality of a new accommodative intraocular lens: An experimental study in rabbits

期刊

HELIYON
卷 9, 期 9, 页码 -

出版社

CELL PRESS
DOI: 10.1016/j.heliyon.2023.e19604

关键词

Presbyopia; Accommodative lens; Intraocular lens; Preclinical test; Biocompatibility

向作者/读者索取更多资源

This study evaluates the safety, biocompatibility, and functionality of a new accommodative intraocular lens (IOL) in rabbit eyes. The results show that the new lens remained centered and had fewer complications compared to the control group. Functional studies revealed no significant differences between the two materials. Overall, the new IOL showed minimal local biological effects and is comparable to the control group.
This study aims to evaluate the safety, biocompatibility, and functionality of a new accommodative intraocular lens (IOL) (LUZ, patent PCT/ES2016/070,813) after implantation in rabbit eyes. LUZ (Study) and EyeCee & REG; plus a capsular ring (Control) were implanted in rabbits (n = 8 each) after phacoemulsification. Intraoperative follow-up, long-term clinical follow-up, and functional IOL studies were carried out periodically for up to 180 days. A macroscopic examination of the eyeballs to reveal abnormalities and determine the implant centering and a microscopic examination to semi-quantify cell and tissue response were performed. Statistical analysis of the collected data was finally achieved. During follow-up, no significant changes in the general condition nor the clinical evaluation were observed between both groups. However, Study IOL remained centered throughout the study and did not present severe complications as observed in the Control group. Functional studies did not reveal significant differences between both materials. Study showed better centering, fewer adhesions, and maintenance of an opening capsular bag compared to the Control. Local biological effects caused by Study implantation are minimal and comparable to the Control. Therefore, LUZ showed no clinical signs or histological response of adverse reaction to the implanted material, according to UNE-EN ISO 11979-5 and 10993-6. Functionality must be confirmed in another animal species with greater lens accommodation capacity than the rabbit. LUZ keeps the capsular bag open, favoring its centering and avoiding fibrosis and adherence to the bag; this allows potential accommodation of this IOL and theoretically enables the patient to focus dynamically.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据