期刊
PHOTONICS
卷 10, 期 9, 页码 -出版社
MDPI
DOI: 10.3390/photonics10090968
关键词
X-ray; phase-contrast imaging; gratings; single-shot
类别
In this paper, a single grating imaging system is proposed for X-ray phase-contrast imaging. The system uses a single-shot algorithm to retrieve attenuation and phase gradient distribution, achieving time-saving and detail preservation. By using a large-size X-ray detector, the system can be configured for both microscopic and large field-of-view imaging.
In the field of X-ray phase-contrast imaging, a time-saving approach and preservation of details are crucial factors for obtaining phase-contrast images. In this manuscript, a single grating imaging system is proposed to perform the X-ray phase-contrast imaging. Instead of the time-consuming phase-stepping method, this system uses a single-shot algorithm to retrieve the distribution of samples' attenuation and phase gradient. Unlike the single-shot Fourier transform algorithm, which truncates the high-frequency component of the image and reduces the spatial resolution, our method can retrieve the attenuation and phase information images with the same spatial resolution as the images acquired directly by the X-ray detector used. Furthermore, by using a large-size X-ray detector (29 cm x 23 cm), the imaging system can be configured as either a microscopic instrument or a normal large field-of-view imaging system. Finally, a series of experiments were performed to validate the feasibility of the proposed method.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据