This retrospective study examines the effectiveness and safety of a dedicated ECCO2R device in critically ill patients. The results show that the device is able to effectively remove carbon dioxide, improve gas exchange, and has few side effects.
Background: In this retrospective study, we report the effectiveness and safety of a dedicated extracorporeal carbon dioxide removal (ECCO2R) device in critically ill patients. Methods: Adult patients on mechanical ventilation due to acute respiratory distress syndrome (ARDS) or decompensated chronic obstructive pulmonary disease (dCOPD), who were treated with a dedicated ECCO2R device (CO2RESET, Eurosets, Medolla, Italy) in case of hypercapnic acidemia, were included. Repeated measurements of CO2 removal (VCO2) at baseline and 1, 12, and 24 h after the initiation of therapy were recorded. Results: Over a three-year period, 11 patients received ECCO2R (median age 60 [43-72] years) 3 (2-39) days after ICU admission; nine patients had ARDS and two had dCOPD. Median baseline pH and PaCO2 levels were 7.27 (7.12-7.33) and 65 (50-84) mmHg, respectively. With a median ECCO2R blood flow of 800 (500-800) mL/min and maximum gas flow of 6 (2-14) L/min, the VCO2 at 12 h after ECCO2R initiation was 157 (58-183) mL/min. Tidal volume, respiratory rate, and driving pressure were significantly reduced over time. Few side effects were reported. Conclusions: In this study, a dedicated ECCO2R device provided a high VCO2 with a favorable risk profile.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据