Robotic transcranial magnetic stimulation in the treatment of depression: a pilot study

期刊

SCIENTIFIC REPORTS
卷 13, 期 1, 页码 -

出版社

NATURE PORTFOLIO
DOI: 10.1038/s41598-023-41044-1

关键词

-

向作者/读者索取更多资源

A custom-TMS robot was developed for better TMS coil placement, which reduced the rTMS preparation time by 53% and significantly decreased position and orientation errors compared to the manual method. Clinical and neuroimaging assessments indicated comparable improvements in depression severity and regional cerebral blood flow between the robotic and manual rTMS groups.
There has been an increasing demand for robotic coil positioning during repetitive transcranial magnetic stimulation (rTMS) treatment. Accurate coil positioning is crucial because rTMS generally targets specific brain regions for both research and clinical application with other reasons such as safety, consistency and reliability and individual variablity. Some previous studies have employed industrial robots or co-robots and showed they can more precisely stimulate the target cortical regions than traditional manual methods. In this study, we not only developed a custom-TMS robot for better TMS coil placement but also analyzed the therapeutic effects on depression. Treatment effects were evaluated by measuring regional cerebral blood flow (rCBF) using single-photon emission computed tomography and depression severity before and after rTMS for the two positioning methods. The rTMS preparation time with our robotic coil placement was reduced by 53% compared with that of the manual method. The position and orientation errors were also significantly reduced from 11.17 mm and 4.06 degrees to 0.94 mm and 0.11 degrees, respectively, confirming the superiority of robotic positioning. The results from clinical and neuroimaging assessments indicated comparable improvements in depression severity and rCBF in the left dorsolateral prefrontal cortex between the robotic and manual rTMS groups. A questionnaire was used to determine the patients' feelings about the robotic system, including the safety and preparation time. A high safety score indicated good acceptability of robotic rTMS at the clinical site.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据