4.3 Article

Diabetic ketoacidosis due to a sensor defect of FreeStyle Libre: A case report

期刊

出版社

WILEY
DOI: 10.1111/jdi.14065

关键词

Diabetic ketoacidosis; Intermittently scanned continuous glucose monitoring; Type 1 diabetes

向作者/读者索取更多资源

The FreeStyle Libre Flash Glucose Monitoring System can provide sensor glucose values through scanning. However, this case report highlights the importance of considering sensor defects as they can lead to severe complications like diabetic ketoacidosis. Even though the device showed lower glucose values, the patient mistook it for hypoglycemia and reduced insulin dosage, ultimately leading to diabetic ketoacidosis. It is crucial to remind patients about the potential risk of sensor defects despite the benefits of glucose monitoring with this system.
The FreeStyle Libre Flash Glucose Monitoring System allows users to obtain sensor glucose values by scanning with the reader or their mobile phone. We report a case of a 59-year-old man with type 1 diabetes mellitus who developed diabetic ketoacidosis due to a sensor defect. After replacing the sensor with a new one, the glucose value shown in the device was much lower than usual, which made him consider that he was hypoglycemic. Accordingly, he reduced his insulin dose and eventually developed diabetic ketoacidosis. He was unaware of the discrepancy due to the lack of self-monitoring of his blood glucose, although he was educated to do. In sum, glucose monitoring with the FreeStyle Libre is helpful; however, it is necessary to remind the patient that a sensor defect leading to a severe complication frequently happens.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据