期刊
BIOMEDICAL OPTICS EXPRESS
卷 14, 期 8, 页码 4112-4125出版社
Optica Publishing Group
DOI: 10.1364/BOE.491611
关键词
-
This study proposes a novel scanning protocol called ammonite scan for widefield optical coherence tomography angiography (OCTA) and relative retinal blood flow velocity imaging in the human retina using variable interscan time analysis (VISTA). The protocol achieved an interscan time of 1.28 ms through a repeated circle scan with a 400 kHz swept-source. The study also eliminated image artifacts caused by eye movement through extra motion-correction processing and explored the microvascular flow parameter in the healthy human eye using the VISTA image.
A novel scanning protocol, ammonite scan, is proposed for widefield optical coherence tomography angiography (OCTA) and relative retinal blood flow velocity imaging in the human retina using variable interscan time analysis (VISTA). A repeated circle scan using a 400 kHz swept-source was employed to achieve an interscan time of 1.28 ms. The center of the repeated circular scan continuously moved spirally towards the peripheral region, ensuring an extended and adjustable scan range while preserving the short interscan time. Image artifacts due to eye movement were eliminated via extra motion-correction processing using data redundancy. The relative blood flow velocity in superficial and deep plexus layers was calculated from the VISTA image, and their ratio was used to explore the microvascular flow parameter in the healthy human eye.& COPY; 2023 Optica Publishing Group under the terms of the Optica Open Access Publishing Agreement
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐