期刊
SENSORS
卷 23, 期 17, 页码 -出版社
MDPI
DOI: 10.3390/s23177597
关键词
stress phase angle; optical coherence tomography; circumferential stress; wall shear stress
This article proposes a novel method for measuring the stress phase angle (SPA) using spectral domain optical coherence tomography (SD-OCT). The method is validated by measuring SPA in the outflow tract (OFT) of chick embryonic hearts and the common carotid artery of mice. This is the first time that OCT has been used for SPA measurement.
The stress phase angle (SPA), defined as the temporal phase angle between circumferential stress (CS) in the arterial wall and wall shear stress (WSS), is utilized to investigate the interactions between CS and WSS. SPA serves as an important parameter for the early diagnosis of cardiovascular disease. In this study, we proposed a novel method for measuring SPA using spectral domain optical coherence tomography (SD-OCT). The multi-M-mode scan strategy is adopted for interference spectrum acquisition. The phases of CS and WSS are extracted from the corresponding structural and flow velocity images of SD-OCT. The method is validated by measuring SPA in the outflow tract (OFT) of chick embryonic hearts and the common carotid artery of mice. To the best of our knowledge, this is the first time that OCT has been used for SPA measurement.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐