Dose-response dependencies of OSL dosimeters in conventional linacs and 1.5T MR-linacs: an experimental and Monte Carlo study

期刊

PHYSICS IN MEDICINE AND BIOLOGY
卷 68, 期 22, 页码 -

出版社

IOP Publishing Ltd
DOI: 10.1088/1361-6560/ad051e

关键词

optically stimulated luminescence; dosimetry; OSLD; MR-linac; radiotherapy; quality assurance

向作者/读者索取更多资源

This study focuses on the dose-response characterization of optically stimulated luminescence dosimetry (OSLD), with emphasis on 1.5T MR-Linacs. The results show that the dose-response supralinearity in an MR-Linac setting agrees with that in a conventional linac. The signal fading rate of OSLD does not depend on irradiation conditions. The angular dependence of OSLD is more pronounced under the presence of a magnetic field. OSLDs irradiated with and without real-time MR imaging yield the same response. Correction factors for different detector orientations were determined.
Objective. This work focuses on the optically stimulated luminescence dosimetry (OSLD) dose-response characterization, with emphasis on 1.5T MR-Linacs. Approach. Throughout this study, the nanoDots OSLDs (Landauer, USA) were considered. In groups of three, the mean OSLD response was measured in a conventional linac and an MR-Linac under various irradiation conditions to investigate (i) dose-response linearity with and without the 1.5T magnetic field, (ii) signal fading rate and its dependencies, (iii) beam quality, detector orientation and dose rate dependencies in a conventional linac, (iii) potential MR imaging related effects on OSLD response and (iv) detector orientation dependence in an MR-Linac. Monte Carlo calculations were performed to further quantify angular dependence after rotating the detector around its central axis parallel to the magnetic field, and determine the magnetic field correction factors, kB,Q, for all cardinal detector orientations. Main results. OSLD dose-response supralinearity in an MR-Linac setting was found to agree within uncertainties with the corresponding one in a conventional linac, for the axial detector orientation investigated. Signal fading rate does not depend on irradiation conditions for the range of 3-30 d considered. OSLD angular (orientation) dependence is more pronounced under the presence of a magnetic field. OSLDs irradiated with and without real-time T2w MR imaging enabled during irradiation yielded the same response within uncertainties. kB,Q values were determined for all three cardinal orientations. Corrections needed reached up to 6.4%. However, if OSLDs are calibrated in the axial orientation and then irradiated in an MR-Linac placed again in the axial orientation (perpendicular to the magnetic field), then simulations suggest that kB,Q can be considered unity within uncertainties, irrespective of the incident beam angle. Significance. This work contributes towards OSLD dose-response characterization and relevant correction factors availability. OSLDs are suitable for QA checks in MR-based beam gating applications and in vivo dosimetry in MR-Linacs.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据