4.6 Article

Tri-directional x-ray phase contrast multimodal imaging using one hexagonal mesh modulator

期刊

PHYSICS IN MEDICINE AND BIOLOGY
卷 68, 期 19, 页码 -

出版社

IOP Publishing Ltd
DOI: 10.1088/1361-6560/acf5c3

关键词

X-ray phase contrast imaging; structured illumination; phase retrieval; image reconstruction; Fourier transform

向作者/读者索取更多资源

X-ray phase contrast imaging is a promising technique that provides enhanced contrast in soft tissues compared to traditional x-ray attenuation-contrast imaging. This study proposes a mesh-based x-ray phase contrast imaging method with one hexagonal mesh for easy alignment and better image visualization. The hexagonal mesh outperforms the traditional square mesh in image evaluation metrics performance and false color visualization with the same radiation dose, showing its feasibility in real imaging systems and advantages in quantitive imaging and better visualization.
Objective. X-ray phase contrast imaging is a promising technique for future clinical diagnostic as it can provide enhanced contrast in soft tissues compared to traditional x-ray attenuation-contrast imaging. However, the strict requirements on the x-ray coherence and the precise alignment of optical elements limit its applications towards clinical use. To solve this problem, mesh-based x-ray phase contrast imaging method with one hexagonal mesh is proposed for easy alignment and better image visualization. Approach. The mesh produces structured illuminations and the detector captures its distortions to reconstruct the absorption, differential phase contrast (DPC) and dark-field (DF) images of the sample. In this work, we fabricated a hexagonal mesh to simultaneously retrieve DPC and DF signals in three different directions with single shot. A phase retrieval algorithm to obtain artifacts-free phase from DPC images with three different directions is put forward and false color dark-field image is also reconstructed with tri-directional images. Mesh-shifting method based on this hexagonal mesh modulator is also proposed to reconstruct images with better image quality at the expense of increased dose. Main results. In numerical simulations, the proposed hexagonal mesh outperforms the traditional square mesh in image evaluation metrics performance and false color visualization with the same radiation dose. The experimental results demonstrate its feasiblity in real imaging systems and its advantages in quantitive imaging and better visualization. The proposed hexagonal mesh is easy to fabricate and can be successfully applied to x-ray source with it spot size up to 300 & mu;m. Significance. This work opens new possibilities for quantitative x-ray non-destructive imaging and may also be instructive for research fields such as x-ray structured illumination microscopy (SIM), x-ray spectral imaging and x-ray phase contrast and dark-field computed tomography (CT).

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据