4.4 Article

A method for evaluating the force factor in oscillating Kibble balances with application to the Planck-Balance

期刊

METROLOGIA
卷 60, 期 5, 页码 -

出版社

IOP Publishing Ltd
DOI: 10.1088/1681-7575/acef23

关键词

Kibble balance; Planck-Balance; force factor; geometric factor; mass metrology

向作者/读者索取更多资源

This paper presents a method for using high harmonics correction in a table-top version of a Kibble balance, allowing it to be used as a conventional balance in a short motion range.
The Planck-Balance is a table-top version of a Kibble balance, developed in a collaboration between the PTB and TU Ilmenau. In contrast to most Kibble balances, where the motion range is of up to several centimeters in the velocity mode, in the Planck-Balance it is below 100 mu m. For a reliable determination of the force factor, the voice coil is set into harmonic oscillation, resulting in an induced AC voltage. The amplitude of the first harmonic is then taken to determine the force factor. However, due to the non-linear magnetic field across the motion range, the first harmonic is subject to a bias, and higher harmonics are excited. Those higher harmonics contain the information about the relative magnitude of the bias in the first harmonic, and can be used for a correction. In this paper we present an analytical model and show its application with real measurements. The estimated relative bias amounts to -6.3 center dot 10(-6), with a residual relative uncertainty of 6.3 center dot 10(-7), for an oscillation amplitude of 20 mu m. This method is especially of importance for Kibble balances with short motion range and working with harmonic oscillation. After such a correction is applied, a table-top version of a Kibble balance can be used as a conventional balance calibrated by means of E2 mass standards in the whole E2 class mass range.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据