The Effect of Defective PET Detectors in Clinical Simultaneous [18F]FDG Time-of-Flight PET/MR Imaging

期刊

MOLECULAR IMAGING AND BIOLOGY
卷 19, 期 4, 页码 626-635

出版社

SPRINGER
DOI: 10.1007/s11307-016-1023-0

关键词

Defective detectors; Time-of-flight; PET/MR; Clinical; Simulation

资金

  1. Bayer Healthcare
  2. Siemens Healthcare
  3. Roche Pharmaceuticals
  4. GE Healthcare

向作者/读者索取更多资源

The purpose of this study was to evaluate the effect of defective positron emission tomography (PET) detectors on clinical PET image quality in simultaneous PET/magnetic resonance imaging (MRI) for both time-of-flight (TOF) and non-TOF reconstructed images. A total of six patients with various malignant tumors were included and underwent a 2-deoxy-2-[F-18]fluoro-d-glucose PET scan in a fully functional simultaneous TOF PET/MRI. TOF and non-TOF PET images were reconstructed before and after simulating defective detector units. All images were clinically assessed and scored. In addition, a quantitative assessment was performed. Differences were ascertained and compared using the Wilcoxon matched pairs signed-rank test. Without TOF, the image artifacts introduced by one defective detector unit already started to degrade the overall image quality. It reduced the confidence and could lead to a change in diagnosis. Simulating three or five defective detector units resulted in more artifacts and further reduced overall image quality and confidence. By including TOF information, the effects were mitigated: Images reconstructed with one defective detector unit had similar scores as the ones without defective units. The average absolute percentage error for one, three, and five defective detector units were respectively 8, 20, and 37 % for the non-TOF cases and only 5, 11, and 19 % for the TOF cases. Our study indicates that PET image artifacts due to (simulated) defective detectors are significantly mitigated with the integration of TOF information in simultaneous PET/MR. One defective detector unit introduces, on average, a 5 % absolute percentage error. However, in TOF imaging, even in cases with one or three defective units for head and neck imaging and one defective unit for chest and abdominal imaging, overall image quality, artifact scoring, and reader confidence are not significantly degraded.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据