4.5 Article

A temperature-controlled cooling system for accurate quantitative post-mortem MRI

期刊

MAGNETIC RESONANCE IN MEDICINE
卷 -, 期 -, 页码 -

出版社

WILEY
DOI: 10.1002/mrm.29816

关键词

MRI; post-mortem; quantitative imaging; temperature; unfixed

向作者/读者索取更多资源

A temperature-controlled cooling system was developed to facilitate accurate quantitative post-mortem MRI and enable scanning of unfixed tissue. The system was optimized for operational convenience and rapid deployment to ensure efficient workflow. The results showed that the developed cooling system stabilized tissue temperature and enabled accurate estimation of biophysical model parameters, while the conventional setup showed biased estimation.
PurposeTo develop a temperature-controlled cooling system to facilitate accurate quantitative post-mortem MRI and enable scanning of unfixed tissue. MethodsA water cooling system was built and integrated with a 7T scanner to minimize temperature drift during MRI scans. The system was optimized for operational convenience and rapid deployment to ensure efficient workflow, which is critical for scanning unfixed post-mortem samples. The performance of the system was evaluated using a 7-h diffusion MRI protocol at 7T with a porcine tissue sample. Quantitative T-1, T-2, and ADC maps were interspersed with the diffusion scans at seven different time points to investigate the temperature dependence of MRI tissue parameters. The impact of temperature changes on biophysical model fitting of diffusion MRI data was investigated using simulation. ResultsTissue T-1, T-2, and ADC values remained stable throughout the diffusion MRI scan using the developed cooling system, but varied substantially using a conventional scan setup without temperature control. The cooling system enabled accurate estimation of biophysical model parameters by stabilizing the tissue temperature throughout the diffusion scan, while the conventional setup showed evidence of significantly biased estimation. ConclusionA temperature-controlled cooling system was developed to tackle the challenge of heating in post-mortem imaging, which shows potential to improve the accuracy and reliability of quantitative post-mortem imaging and enables long scans of unfixed tissue.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据