Referencing the center of the femoral head during robotic or computer-navigated primary total knee arthroplasty results in less femoral component flexion than the traditional intramedullary axis

期刊

KNEE
卷 44, 期 -, 页码 172-179

出版社

ELSEVIER
DOI: 10.1016/j.knee.2023.08.006

关键词

Femoral component flexion; Computer-assisted TKA; Robotic TKA; Total knee arthroplasty; Knee kinematics

向作者/读者索取更多资源

During robotic and computer-navigated primary total knee arthroplasty, using the center of the femoral head as the reference point for femoral component position results in 1.4 degrees less flexion compared with referencing the intramedullary canal. Surgeons should be aware of these differences as they may ultimately influence knee kinematics.
Background: During robotic and computer-navigated primary total knee arthroplasty (TKA), the center of the femoral head is utilized as the proximal reference point for femoral component position rather than the intramedullary axis. We sought to analyze the effect on femoral component flexion-extension position between these two reference points.Methods: We obtained CT 3D-reconstructions of 50 cadaveric intact femurs. We defined the navigation axis as the line from center of the femoral head to center of the knee (lowest point of the trochlear groove) and the intramedullary axis as the line from center of the knee to center of the canal at the isthmus. Differences between these axes in the sagittal plane were measured. Degree of femoral bow and femoral neck anteversion were correlated with the differences between the two femoral axes.Results: On average, the navigated axis was 1.4 degrees (range, -1.4 degrees to 4.1 degrees) posterior to the intramedullary axis. As such, the femoral component would have on average 1.4 degrees less flexion compared with techniques referencing the intramedullary canal. A more anterior intramedullary compared with navigated axis (i.e., less femoral flexion) was associated with more femoral bow (R-2 = 0.7, P < 0.001) and less femoral neck anteversion (R-2 = 0.5, P < 0.05).Conclusion: Computer-navigated or robotic TKA in which the center of the femoral head is utilized as a reference point, results in 1.4 degrees less femoral component flexion than would be achieved by referencing the intramedullary canal. Surgeons should be aware of these differences as they may ultimately influence knee kinematics.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据