4.6 Article

Electrochemical Determination of Paraquat Using Ordered Mesoporous Carbon (CMK-3) Modified Glassy Carbon Electrode

期刊

出版社

ELECTROCHEMICAL SOC INC
DOI: 10.1149/1945-7111/acedd0

关键词

paraquat; CMK-3; GCE; anodic stripping squarewave voltammetry; limit of detection; vegetable and fruit extracts

向作者/读者索取更多资源

Ordered mesoporous carbon CMK-3 was synthesized and characterized using various physicochemical and electroanalytical tools. The modified CMK-3 electrode exhibited enhanced sensitivity for the detection of herbicide paraquat compared to the bare electrode. The electrode showed a linear variation in sensitivity with concentration and the detection process was diffusion-controlled.
Ordered mesoporous carbon namely, CMK-3 was synthesized by nano casting method, and characterized using various physicochemical and electroanalytical tools like HESEM, HRTEM, BET, etc. Using Cyclic voltammetric (CV) measurements on CMK-3 modified glassy carbon electrode (CMK-3/GCE) in the standard redox couple, [Fe(CN)(6)](3-/4-), the electrochemical surface area was deduced as 0.081 cm(2). The modified electrode was used as an electrocatalyst for the determination of one of the herbicides, paraquat (PQ). The modified electrode shows enhanced sensitivity for the determination of PQ in phosphate buffer solution at pH 9.2 as compared to bare GCE. The concentration dependant CV studies showed a linear variation in sensitivity in the concentration range of 60 & mu;M to 450 & mu;M. The effect of the scan rate shows a linear variation in sensitivity as a function of the square root of scan rate suggesting a diffusion-controlled process during the detection of PQ. Anodic stripping square wave voltammetry (ASSWV) reveals the sensitivity and limit of detection as 9.51 & mu;A & mu;M-1 and 64 nM, respectively. The interfering ability of the other foreign molecules like metal ions, environmental samples, biomolecules, and pesticides was analyzed using CMK-3/GCE. The results revealed that the electrode is able to sense paraquat, with the retention in sensitivity ranges between 91% and 102%. The real-time applicability of the electrode was examined using vegetable and fruit extracts. The analysis revealed that the recovery values were between 102 % - 105 %.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据