Active drug-coated flow diverter in a preclinical model of intracranial stenting

期刊

出版社

BMJ PUBLISHING GROUP
DOI: 10.1136/jnis-2023-020391

关键词

flow diverter; aneurysm; bioactive; device; MRI

向作者/读者索取更多资源

This study tested the effectiveness of a heparin-coated flow diverter in reducing the risk of thromboembolic complications. The results showed that the coated device significantly reduced the number of new magnetic susceptibility artifacts after 1 week follow-up, indicating its potential in reducing thromboembolic complications.
BackgroundFlow diverters carry the risk of thromboembolic complications (TEC). We tested a coating with covalently bound heparin that activates antithrombin to address TEC by locally downregulating the coagulation cascade. We hypothesized that the neuroimaging evidence of TEC would be reduced by the coating. Methods16 dogs were implanted with overlapping flow diverters in the basilar artery, separated into two groups: heparin-coated (n=9) and uncoated (n=7). Following implantation, high-frequency optical coherence tomography (HF-OCT) was acquired to quantify acute thrombus (AT) formation on the flow diverters. MRI was performed postoperatively and repeated at 1, 2, 3, 4, and 8 weeks, consisting of T1-weighted imaging, time-0f-flight (ToF), diffusion weighted imaging (DWI), susceptibility weighted imaging (SWI), and fluid attenuated inversion recovery (FLAIR) sequences. Neurological examinations were performed throughout the 8-week duration of the study. ResultsThe mean AT volume on coated devices was lower than uncoated (0.014 vs 0.018 mm(3)); however, this was not significant (P=0.3). The mean number of foci of magnetic susceptibility artifacts (MSAs) on SWI was significantly different between the uncoated and coated groups at the 1-week follow-up (P<0.02), and remained statistically different throughout the duration of the study. The AT volume showed a direct linear correlation with the MSA count and 80% of the variance in the MSA could be explained by the AT volume (P<0.001). Pathological analysis showed evidence of ischemic injury at locations of MSA. ConclusionsHeparin-coated flow diverters significantly reduced the number of new MSAs after 1 week follow-up, showing the potential to reduce TEC.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据