Validation of a multiple-lead smartphone-based electrocardiograph with automated lead placement for layman use in patients with hypertrophic cardiomyopathy

期刊

JOURNAL OF ELECTROCARDIOLOGY
卷 79, 期 -, 页码 1-7

出版社

CHURCHILL LIVINGSTONE INC MEDICAL PUBLISHERS
DOI: 10.1016/j.jelectrocard.2023.02.006

关键词

Smartphone ECG; Layman; Screening; Hypertrophic cardiomyopathy

向作者/读者索取更多资源

This study validated the accuracy of the D-Heart ECG device, a smartphone 8/12-Lead electrocardiograph with an image processing algorithm, for diagnosing ECG abnormalities in patients with hypertrophic cardiomyopathy. The results showed that the D-Heart ECG device had high accuracy and concordance with standard 12-Lead ECG, and the image processing algorithm provided accurate electrode placement, standardizing the quality of the examination.
Background: A smartphone 12-Lead ECG that enables layman ECG screening is still lacking. We aimed to validate D-Heart ECG device, a smartphone 8/12 Lead electrocardiograph with an image processing algorithm to guide secure electrode placement by non-professional users. Methods: One-hundred-fourty-five patients with HCM were enrolled. Two uncovered chest images were acquired using the smartphone camera. An image with virtual electrodes placement by imaging processing algorithm software was compared to the 'gold standard' electrode placement by a doctor. D-Heart 8 and 12-Lead ECG were obtained, immediately followed by 12-lead ECGs and were assessed by 2 independent observers. Burden of ECG abnormalities was defined by a score based on the sum of 9 criteria, identifying four classes of increasing severity. Results: A total of 87(60%) patients presented a normal/mildly abnormal ECG, whereas 58(40%) had moderate or severe ECG alteration. Eight(6%) patients had >= 1 misplaced electrode. D-Heart 8-Lead and 12-lead ECGs concordance according to Cohen's weighted kappa test was 0,948 (p < 0,001, agreement of 97.93%). Concordance was high for the Romhilt-Estes score (kw = 0,912; p < 0.01). Concordance between D-Heart 12-Lead ECG and standard 12-Lead ECG was perfect (kw = 1). PR and QRS intervals measurements comparison with Bland-Altman method showed good accuracy (95% limit of agreement +/- 18 ms for PR and +/- 9 ms for QRS). Conclusions: D-Heart 8/12-Lead ECGs proved accurate, allowing an assessment of ECG abnormalities comparable to the standard 12-lead ECG in patients with HCM. The image processing algorithm provided accurate electrode placement, standardizing exam quality, potentially opening perspectives for layman ECG screening campaigns.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.1
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据