4.5 Article

Accurate method for value assignment of carcinoembryonic antigen reference materials

期刊

出版社

WILEY
DOI: 10.1002/jcla.24936

关键词

carcinoembryonic antigen; commutability; comparability; reference material

向作者/读者索取更多资源

This study investigated the comparability of reference materials (RMs) for carcinoembryonic antigen (CEA) and selected an appropriate diluent matrix for CEA measurement. The results showed that the candidate RMs for CEA were commutable among all immunoassays, and the diluted WHO 73/601 RMs in PBS buffer matrix were selected as common calibrators for five immunoassays, effectively improving the harmonization of CEA detection.
BackgroundIn this study, we explored the commutability of reference materials (RMs) for carcinoembryonic antigen (CEA), selected the appropriate diluent matrix of the first International Reference Preparation (IRP) 73/601 of the World Health Organization (WHO 73/601) for CEA, and improved the comparability of CEA measurement results among different assay systems. MethodsForty serum samples were divided into five aliquots. WHO 73/601 was diluted into nine concentrations using five diluents with different components, and the candidate RMs for CEA at five concentrations (C1-C5) were prepared by the Beijing Clinical Laboratory Center (BCCL). The samples were analyzed via five automated CEA immunoassays. ResultsCarcinoembryonic antigen candidate RMs were commutable among all immunoassays based on the CLSI approach and among 7 of 10 assay combinations based on the IFCC approach. WHO 73/601 diluted in phosphate-buffered saline (PBS) was commutable among all assays based on the CLSI approach and among 5 of 10 pairwise comparisons based on the IFCC approach with correction of bias at diluted concentrations, except for the lowest concentration, which had the smallest variation among systems. The median percentage biases among assays were decreased after calibration. ConclusionThe BCCL candidate RMs (C2-C5) for CEA were commutable among all immunoassays. WHO 73/601 RMs diluted in a PBS buffer matrix were selected as common calibrators for five immunoassays, which reduced bias, thereby effectively improving the harmonization of CEA detection; therefore, they could be used to assign values to CEA candidate RMs developed by BCCL. Our findings promote the harmonization of CEA detection in immunoassays.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据