Managing sample introduction problems in hydrophilic interaction liquid chromatography

期刊

JOURNAL OF CHROMATOGRAPHY A
卷 1700, 期 -, 页码 -

出版社

ELSEVIER
DOI: 10.1016/j.chroma.2023.464006

关键词

HPLC; HILIC; Sample injection

向作者/读者索取更多资源

Sample injection in HILIC can cause issues when the injection solvent has higher elution strength than the mobile phase, resulting in asymmetric peak shapes and poor efficiency. Strategies to overcome this problem include pre-column dilution with a weak solvent and using alternative organic solvents in the injection solvent. Controlling the volume of injection solvents and using small sample volumes is recommended to reduce mismatch effects.
Sample injection can cause serious problems in hydrophilic interaction liquid chromatography (HILIC) when the injection solvent has higher elution strength than the mobile phase (mp). It can lead to asym-metric peak shapes and poor efficiency. The problem can occur when the mp contains a high proportion of organic e.g. 95% acetonitrile (a weak solvent) whereas the injection solvent contains a higher propor-tion of water (a strong solvent) that is necessary to dissolve polar samples. We investigated different strategies to overcome this problem. A simple method is pre-column dilution where the injector is pro-grammed to deliver a plug of weak solvent (e.g. pure acetonitrile) along with the sample dissolved in a solvent with higher water content than the mp. Another option is to use alternative organic solvents to acetonitrile in the injection solvent, e.g. isopropanol, acetone or tetrahydrofuran, that may give enhanced sample solubility. The role of the volume of injection solvents was investigated as well as the possible ef-fects of mass overload on the results. The use of small sample volumes is always recommended to reduce mismatch effects.(c) 2023 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据