4.5 Article

Impact of leaflet thrombosis on valve haemodynamic status after transcatheter aortic valve replacement

期刊

HEART
卷 -, 期 -, 页码 -

出版社

BMJ PUBLISHING GROUP
DOI: 10.1136/heartjnl-2023-322946

关键词

aortic valve stenosis; transcatheter aortic valve replacement

向作者/读者索取更多资源

In patients who underwent successful transcatheter aortic valve replacement (TAVR), the presence of subclinical leaflet thrombosis did not significantly affect valve hemodynamic status and was not a predictor of structural valve deterioration. Further long-term follow-up is needed to evaluate its impact on valve durability.
Objectives The effect of subclinical leaflet thrombosis, characterised by hypoattenuated leaflet thickening (HALT), on the valve haemodynamic function and durability of the bioprosthetic valve, is not yet determined. We determined the impact of HALT on valve haemodynamics after transcatheter aortic valve replacement (TAVR) and the predictors of haemodynamic structural valve deterioration (SVD). Methods The Anticoagulation vs Dual Antiplatelet Therapy for Prevention of Leaflet Thrombosis and Cerebral Embolization after Transcatheter Aortic Valve Replacement(ADAPT-TAVR) trial is a multicenter, randomised trial that compared edoxaban and dual antiplatelet therapy in patients who had undergone successful TAVR. The presence of HALT was evaluated by four-dimensional CT at 6 months and serial echocardiography performed at baseline, immediately post-TAVR and after 6 months. SVD was defined as at least one of the following: (1) mean transprosthetic gradient >= 20 mm Hg, (2) change in the mean gradient >= 10 mm Hg from baseline, or (3) new or increase in intraprosthetic aortic regurgitation of at least >= 1 grade, resulting in moderate or greater regurgitation. Results At 6 months, HALT was found in 30 of 211 (14.2%) patients. The presence of HALT did not significantly affect aortic valve mean gradients (with vs without HALT; 14.0 +/- 4.8 mm Hg vs 13.7 +/- 5.5 mm Hg; p=0.74) at 6 months. SVD was reported in 30 of 206 patients (14.6%) at 6-month follow-up echocardiography. Older age (OR: 1.138; 95% CI: 1.019 to 1.293; p=0.033), use of aortic valve size <= 23 mm (OR: 6.254; 95% CI: 2.230 to 20.569; p=0.001) and mean post-TAVR pressure gradient (OR: 1.233; 95% CI: 1.123 to 1.371; p<0.001) were independent predictors of haemodynamic SVD; however, the presence of HALT was not identified as a predictor of SVD. Conclusions In patients who had undergone successful TAVR, aortic valve haemodynamic status was not influenced by the presence of HALT. Although HALT was not a predictor of haemodynamic SVD at 6 months, it warrants further longer-term follow-up to evaluate the effect on long-term valve durability.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据