Comparison of corneal measurements in keratoconus eyes with two swept-source-optical coherence tomography devices and a Scheimpflug device

出版社

SPRINGER
DOI: 10.1007/s00417-023-06219-6

关键词

Keratoconus; Eyestar; Pentacam; IOLMaster; Comparison; Keratometry

向作者/读者索取更多资源

The purpose of this study was to assess the feasibility and reliability of biometric measurements taken with the Eyestar 900 device in keratoconus eyes compared to the Pentacam HR and IOLMaster 700 devices. The results showed that the Eyestar 900 demonstrated high repeatability and good correlation with the other devices, particularly in terms of corneal topography and axial length measurements.
Purpose To assess the feasibility and reliability of biometric measurements taken with the Eyestar 900 device in keratoconus eyes in comparison with those taken with the Pentacam HR and IOLMaster 700. Methods Seventy-five eyes of 75 patients with keratoconus were included. The central corneal thickness (CCT), thinnest point of corneal thickness (TCT), axial length (AL), flat (K1) and steep (K2) anterior and posterior (Kp1, Kp2) keratometry, maximal keratometry (KMax) and anterior chamber depth (ACD) were compared between the Eyestar 900, Pentacam HR and IOLMaster 700. Reliability parameters such as the coefficient of variation (CoV) and intraclass correlation coefficient (ICC) were calculated. Pearson's r was determined to assess the correlation between devices. Results A high repeatability (CoV < 1%) and intraclass correlation (ICC > 0.9) was found for all devices, led by AL, TCT, K1 and K2 (CoV 0.01-0.36%; ICC 0.994-1.00). The largest correlation between devices was found for AL ( Eyestar vs. IOLMaster, r = 1.0), K1 (Eyestar vs. IOL-Master, r = 0.997) and ACD (Eyestar vs. IOLMaster, r = 0.995; Pentacam vs. IOL-Master, r = 0.987; Eyestar vs. Pentacam, r = 0.983), but there were significant differences in measured values between devices (p < 0.001), whereas the correlation was only slightly lower (r = 0.947 to 0.994) for KMax, CCT, TCT, K2, Kp1 and Kp2. Conclusion Keratometric and axial length measurements with the Eyestar 900 were feasible and revealed a high repeatability and a good correlation to the other devices in eyes with keratoconus.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据