MEDICAL PHYSICS
卷 43, 期 8, 页码 4833-4841出版社
AMER ASSOC PHYSICISTS MEDICINE AMER INST PHYSICS
DOI: 10.1118/1.4959536
关键词
radiotherapy; patient positioning; surface imaging; marker-less device; time of flight; iterative closest point
资金
- Varian Medical Systems
- Brest Metropole Oceane council
- Plan Cancer [A12147NS]
Purpose: To evaluate the patient positioning accuracy in radiotherapy using a stereo-time of flight (ToF)-camera system. Methods: A system using two ToF cameras was used to scan the surface of the patients in order to position them daily on the treatment couch. The obtained point clouds were registered to (a) detect translations applied to the table (intrafraction motion) and (b) predict the displacement to be applied in order to place the patient in its reference position (interfraction motion). The measures provided by this system were compared to the effectively applied translations. The authors analyzed 150 fractions including lung, pelvis/prostate, and head and neck cancer patients. Results: The authors obtained small absolute errors for displacement detection: 0.8 +/- 0.7, 0.8 +/- 0.7, and 0.7 +/- 0.6 mm along the vertical, longitudinal, and lateral axes, respectively, and 0.8 +/- 0.7 mm for the total norm displacement. Lung cancer patients presented the largest errors with a respective mean of 1.1 +/- 0.9, 0.9 +/- 0.9, and 0.8 +/- 0.7 mm Conclusions: The proposed stereo-ToF system allows for sufficient accuracy and faster patient repositioning in radiotherapy. Its capability to track the complete patient surface in real time could allow, in the future, not only for an accurate positioning but also a real time tracking of any patient intrafraction motion (translation, involuntary, and breathing). (C) 2016 American Association of Physicists in Medicine.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据