4.3 Article

Vie Scope & REG; versus videolaryngoscopy in expected difficult airways: a randomized controlled trial

出版社

SPRINGER
DOI: 10.1007/s12630-023-02534-y

关键词

airway management; diagnostic techniques; laryngoscopy; laryngoscopes; respiratory system; tracheal intubation

向作者/读者索取更多资源

This study compared laryngeal visualization between Vie Scope and Macintosh videolaryngoscopy in patients with an anticipated difficult airway. The results showed that Vie Scope was noninferior to videolaryngoscopy in terms of laryngeal visualization, but the time required for intubation was longer with Vie Scope.
PurposeThe management of patients with an anticipated difficult airway remains challenging. We evaluated laryngeal visualization with the recently introduced Vie Scope & REG; as a straight blade laryngoscope consisting of an illuminated tube necessitating bougie-facilitated intubation vs Macintosh videolaryngoscopy.MethodsWe conducted a prospective randomized controlled noninferiority trial. Patients undergoing elective ear, nose, and throat or oral and maxillofacial surgery with an anticipated difficult airway were randomized 1:1 to receive tracheal intubation with the Vie Scope or Macintosh videolaryngoscope (C-MAC & REG;). The primary outcome measure was laryngeal visualization by the percentage of glottis opening (POGO) scale. Secondary outcome measures were the time to successful intubation (TTI) and first-attempt and overall success rates.ResultsWe included two sets of 29 patients in our analysis. For visualization, the Vie Scope was noninferior to videolaryngoscopy (VL) with mean (standard deviation [SD]) POGO scores of 71 (31)% vs 64 (30)% in the VL group [difference in means, 7 (8)%; 95% confidence interval, -9 to 23; P = 0.38]. Mean (SD) TTI was 125 (129) sec in the Vie Scope and 51 (36) sec in the VL group (difference in means, 75 sec; 95% confidence interval, 25 to 124; P = 0.005). The first-attempt and overall success rates were 22/29 (76%) and 27/29 (93%) in both groups. Two patients per group were switched to a different device. Four accidental esophageal intubations occurred in the Vie Scope group, these were presumably due to bougie misplacement.ConclusionVisualization with the Vie Scope was noninferior to VL in patients with an anticipated difficult airway, but TTI was longer in the Vie Scope group.Study registrationClinicalTrials.gov (NCT05044416); registered 5 September 2021.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据