期刊
MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE
卷 29, 期 1, 页码 75-87出版社
SPRINGER
DOI: 10.1007/s10334-015-0505-4
关键词
Combined PET/MR; MR-based attenuation correction; Whole-body hybrid imaging; Reproducibility; Accuracy
We assess inter- and intra-subject variability of magnetic resonance (MR)-based attenuation maps (MR mu Maps) of human subjects for state-of-the-art positron emission tomography (PET)/MR imaging systems. Four healthy male subjects underwent repeated MR imaging with a Siemens Biograph mMR, Philips Ingenuity TF and GE SIGNA PET/MR system using product-specific MR sequences and image processing algorithms for generating MR mu Maps. Total lung volumes and mean attenuation values in nine thoracic reference regions were calculated. Linear regression was used for comparing lung volumes on MR mu Maps. Intra- and inter-system variability was investigated using a mixed effects model. Intra-system variability was seen for the lung volume of some subjects, (p = 0.29). Mean attenuation values across subjects were significantly different (p < 0.001) due to different segmentations of the trachea. Differences in the attenuation values caused noticeable intra-individual and inter-system differences that translated into a subsequent bias of the corrected PET activity values, as verified by independent simulations. Significant differences of MR mu Maps generated for the same subjects but different PET/MR systems resulted in differences in attenuation correction factors, particularly in the thorax. These differences currently limit the quantitative use of PET/MR in multi-center imaging studies.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐