Comparative Evaluation between Visual and Automated Dipstick Urinalyses in Dogs

期刊

VETERINARY SCIENCES
卷 10, 期 4, 页码 -

出版社

MDPI
DOI: 10.3390/vetsci10040284

关键词

dipstick; dog; urine; automated; visual

向作者/读者索取更多资源

The purpose of this study was to compare visual and automated evaluations of dipstick variables in canine urine samples. Automated analysis of urine dipsticks has improved accuracy and precision. However, comparing the two methods, correlation for urine specific gravity was poor and agreement varied for different variables. Visual and automated dipstick urinalyses should not be used interchangeably, and multiple urine samples from the same dog should be evaluated using the same method for accurate results.
Simple Summary The purpose of the current study was to compare visual and automated evaluations of dipstick variables in canine urine samples. Urine dipstick chemical reaction is based on the color change of a reagent when a specific substance is detected. The automated analysis of urine dipsticks has improved the accuracy and precision of the results. Urine test strips are commercially available and can be assessed with semi-automated analyzers or by visual assessment. This study aimed to compare the visual and automated evaluations of dipstick variables in canine urine samples. One hundred and nineteen urine samples were evaluated. Automated analysis was performed on a veterinary urine analyzer URIT-50Vet (URIT Medical Electronic) with UC VET13 Plus strips. Multistix 10 SG dipsticks (Siemens Healthcare GmbH, Erlangen, Germany) were used for visual evaluation, along with a refractometer (Clinical Refractometer Atago T2-Ne, Atago Co., Tokyo, Japan) for urine specific gravity measurements. A linear relationship was observed between the pH measurements (p = 0.2) of the two methods; the Passing-Bablok procedure was valid since neither proportional nor systematic significant errors were observed. Comparing the two methods, the correlation for urine specific gravity was poor (p = 0.01, CI 0.667-1.000). Moderate agreement was demonstrated for proteins (kappa = 0.431), bilirubin (kappa = 0.434) and glucose (kappa = 0.450). Agreement was substantial for blood (kappa = 0.620) and poor for leukocytes (kappa = 0.100). Poor agreement was observed for ketones (kappa = -0.006). Apart from the pH analysis, visual and automated dipstick urinalyses should not be used interchangeably. Multiple urine samples obtained from the same dog during the day should be evaluated using the same method to overcome erroneous results.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据