Inter-Changeability of Impedance Devices for Lymphedema Assessment

期刊

LYMPHATIC RESEARCH AND BIOLOGY
卷 14, 期 2, 页码 88-94

出版社

MARY ANN LIEBERT, INC
DOI: 10.1089/lrb.2015.0026

关键词

-

向作者/读者索取更多资源

Impedance technology is a popular technique for the early detection of lymphedema. The preferred approach is to use bioimpedance spectroscopy (BIS), with measurements being made with the subject lying supine, although attempts have been made to use single or multiple frequency impedance measurements obtained while the subject is standing. The aim of the present study was to determine the equivalence of these different approaches. Impedance measurements of the individual limbs of 37 healthy individuals were determined using both a stand-on, multi-frequency impedance device and a supine impedance spectroscopy instrument. Significant differences were found between the instruments in both absolute impedance values and, importantly, inter-limb impedance ratios. Since impedance ratios in healthy individuals provide the reference standard for detection of lymphedema, these data indicate that the methods are not interchangeable. Consideration of the errors associated with each method indicates that the BIS remains the preferred method for lymphedema detection.