4.5 Article

Intensive care unit nurses' perceptions and practices regarding clinical alarms: A descriptive study

期刊

NURSING OPEN
卷 10, 期 8, 页码 5531-5540

出版社

WILEY
DOI: 10.1002/nop2.1792

关键词

alarm management; clinical alarm; intensive care unit; nurse; patient safety; physiological monitor

类别

向作者/读者索取更多资源

This study aimed to describe the frequencies of physiologic monitor clinical alarms and investigate nurses' perceptions and practices regarding these alarms in ICUs. The findings revealed that physiologic monitor alarms occur frequently in ICU, and it is recommended to use smart medical devices and alarm notification systems, formulate and implement standardized alarm management policies and norms, and strengthen alarm management education and training to improve nursing quality and patient safety.
Aim: To describe the frequencies of physiologic monitor clinical alarms and to investigate nurses' perceptions and practices regarding clinical alarms in ICUs.Design: A descriptive study.Methods: A 24-h continuous nonparticipant observation study was conducted in ICU. Observers observed and recorded the occurrence time, detail information when electrocardiogram monitor alarms triggered. And a cross-sectional study was conducted among ICU nurses by convenience sampling, using the general information questionnaire and the Chinese version of clinical alarms survey questionnaire for medical devices. Data analysis was performed using SPSS 23.Results: A total of 13,829 physiologic monitor clinical alarms were recorded in 14-day observation and 1191 ICU nurses responded to the survey. Most nurses agreed or strongly agreed the sensitivity to alarms and responded quickly (81.28%), smart alarm systems (74.56%), alarm notification systems (72.04%) and set up alarm administrators (59.45%) were useful to improve alarm management, while frequent nuisance alarms disrupted patients care (62.47%) and reduced nurses' trust in alarms (49.03%), environmental noise interfered with nurses' recognition of the alarms (49.12%) and not everyone received education of alarm systems (64.65%).Conclusions: Physiological monitor alarms occur frequently in ICU, and it is necessary to formulate or further optimize alarm management measures. It is recommended to use smart medical devices and alarm notification systems, formulate and implement standardized alarm management policies and norms, and strengthen alarm management education and training, so as to improve the nursing quality and patient safety.Patient or Public Contribution: The patients in the observation study included all patients admitted to the ICU during the observation period. The nurses in the survey study were conveniently selected through an online survey.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据