Respiratory Rate Estimation during Walking and Running Using Breathing Sounds Recorded with a Microphone

期刊

BIOSENSORS-BASEL
卷 13, 期 6, 页码 -

出版社

MDPI
DOI: 10.3390/bios13060637

关键词

wearable sensors; breathing sounds; validation protocol; respiratory frequency; measurement accuracy; exercise; sport sensors

向作者/读者索取更多资源

Emerging evidence suggests that respiratory frequency is a valid marker of physical effort, leading to the interest in developing devices for monitoring this vital sign. This study proposes the use of a microphone embedded in a facemask to estimate respiratory frequency from breath sounds during walking and running. The results show relatively good agreement between the proposed system and the reference system, indicating that microphone sensors can be considered suitable for estimating respiratory frequency during exercise.
Emerging evidence suggests that respiratory frequency (f(R)) is a valid marker of physical effort. This has stimulated interest in developing devices that allow athletes and exercise practitioners to monitor this vital sign. The numerous technical challenges posed by breathing monitoring in sporting scenarios (e.g., motion artifacts) require careful consideration of the variety of sensors potentially suitable for this purpose. Despite being less prone to motion artifacts than other sensors (e.g., strain sensors), microphone sensors have received limited attention so far. This paper proposes the use of a microphone embedded in a facemask for estimating f(R) from breath sounds during walking and running. f(R) was estimated in the time domain as the time elapsed between consecutive exhalation events retrieved from breathing sounds every 30 s. Data were collected from ten healthy subjects (both males and females) at rest and during walking (at 3 km/h and 6 km/h) and running (at 9 km/h and 12 km/h) activities. The reference respiratory signal was recorded with an orifice flowmeter. The mean absolute error (MAE), the mean of differences (MOD), and the limits of agreements (LOAs) were computed separately for each condition. Relatively good agreement was found between the proposed system and the reference system, with MAE and MOD values increasing with the increase in exercise intensity and ambient noise up to a maximum of 3.8 bpm (breaths per minute) and -2.0 bpm, respectively, during running at 12 km/h. When considering all the conditions together, we found an MAE of 1.7 bpm and an MOD & PLUSMN; LOAs of -0.24 & PLUSMN; 5.07 bpm. These findings suggest that microphone sensors can be considered among the suitable options for estimating f(R) during exercise.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据