4.6 Article

Clinical Performance Evaluation of a Rapid Real-Time PCR Assay for Monkeypox Diagnosis: a Retrospective and Comparative Study

期刊

MICROBIOLOGY SPECTRUM
卷 11, 期 3, 页码 -

出版社

AMER SOC MICROBIOLOGY
DOI: 10.1128/spectrum.00510-23

关键词

clinical validation; molecular diagnostics; monkeypox virus; real-time PCR

向作者/读者索取更多资源

In an increasingly globalized and interconnected world, the outbreak of infectious diseases can become a global health emergency. The current example is the 2022 monkeypox virus outbreak. Strategies to identify and control the spread of the virus should be developed to prevent further crises.
In an increasingly globalized and interconnected world, the outbreak of an infectious disease in one country can become a worrying health emergency for the whole world. A current example is the 2022 monkeypox virus (mpox) outbreak affecting multiple areas across the world. In this context, strategies to interrupt transmission as soon as possible by identifying cases, clusters, and sources of infection should be developed around the world to prevent these crises. The aim of this retrospective and collaborative study was to perform external clinical validation of the VIASURE monkeypox virus real-time PCR detection kit (CerTest Biotec, Spain) with ready-to-use reagents designed for the rapid detection of mpox. A total of 165 samples with suspected infection were used for this analysis. The standard procedures of the clinical microbiology laboratory of the Miguel Servet University Hospital, using the RealStar Orthopoxvirus PCR kit v1.0 (Altona Diagnostics) and bidirectional Sanger sequencing (STAB VIDA, Caparica, Portugal), were considered reference techniques. Furthermore, a subset of 67 mpox-negative samples and 13 mpox-positive samples were routinely tested for clinical diagnosis of other rash/ulcerative pathologies. Accuracy testing resulted in appropriate clinical validation values, as follows: sensitivity, 1 (95% confidence interval [CI], 0.97 to 1); specificity, 1 (95% CI, 0.98 to 1); positive predictive value, 1 (95% CI, 0.93 to 1); negative predictive value, 1 (95% CI, 0.95 to 1). The strength of agreement between assays was almost perfect. The added value is the useful support for the specific diagnosis of mpox infections due to the diagnostic specificity data obtained.IMPORTANCE Given that a large number of mpox outbreaks have been reported worldwide since 2022 in countries in which the disease is not endemic, the main concern for clinicians and global health systems should be to develop effective, available, and easy-to-implement diagnostic strategies to interrupt mpox transmission as soon as possible. This retrospective study demonstrates the satisfactory clinical parameters of a commercially available molecular diagnostic kit for routine testing for mpox in clinical diagnostic laboratories. Given that a large number of mpox outbreaks have been reported worldwide since 2022 in countries in which the disease is not endemic, the main concern for clinicians and global health systems should be to develop effective, available, and easy-to-implement diagnostic strategies to interrupt mpox transmission as soon as possible. This retrospective study demonstrates the satisfactory clinical parameters of a commercially available molecular diagnostic kit for routine testing for mpox in clinical diagnostic laboratories.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据