Breath-by-breath measurement of exhaled ammonia by acetone-modifier positive photoionization ion mobility spectrometry via online dilution and purging sampling

期刊

JOURNAL OF PHARMACEUTICAL ANALYSIS
卷 13, 期 4, 页码 412-420

出版社

ELSEVIER
DOI: 10.1016/j.jpha.2023.02.008

关键词

Photoionization ion mobility spectrometry; Acetone modifier; Exhaled ammonia; Breath analysis

向作者/读者索取更多资源

In this study, an acetone-modifier positive photoionization ion mobility spectrometry (AM-PIMS) method was developed for accurate qualitative and quantitative analysis of exhaled ammonia (NH3) with high selectivity and sensitivity. The introduction of acetone as a modifier improved the peak-to-peak resolution and the accuracy of NH3 qualitative identification. Online dilution and purging sampling reduced interference and memory effects, enabling breath-by-breath measurement. The AM-PIMS method showed great potential for clinical disease diagnosis.
Exhaled ammonia (NH3) is an essential noninvasive biomarker for disease diagnosis. In this study, an acetone-modifier positive photoionization ion mobility spectrometry (AM-PIMS) method was developed for accurate qualitative and quantitative analysis of exhaled NH3 with high selectivity and sensitivity. Acetone was introduced into the drift tube along with the drift gas as a modifier, and the characteristic NH3 product ion peak of (C3H6O)4NH4 thorn (K0 = 1.45 cm2/V center dot s) was obtained through the ion-molecule reaction with acetone reactant ions (C3H6O)2H thorn (K0 = 1.87 cm2/V center dot s), which significantly increased the peak-to-peak resolution and improved the accuracy of exhaled NH3 qualitative identification. Moreover, the interference of high humidity and the memory effect of NH3 molecules were significantly reduced via online dilution and purging sampling, thus realizing breath-by-breath measurement. As a result, a wide quantitative range of 5.87-140.92 mmol/L with a response time of 40 ms was achieved, and the exhaled NH3 profile could be synchronized with the concentration curve of exhaled CO2. Finally, the analytical capacity of AM-PIMS was demonstrated by measuring the exhaled NH3 of healthy subjects, demonstrating its great potential for clinical disease diagnosis. (c) 2023 The Author(s). Published by Elsevier B.V. on behalf of Xi'an Jiaotong University. This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据