Validation of a novel automated system, Fluispotter (R), for serial sampling of dried blood spots

期刊

ENDOCRINE CONNECTIONS
卷 12, 期 7, 页码 -

出版社

BIOSCIENTIFICA LTD
DOI: 10.1530/EC-23-0087

关键词

dried blood spots; cortisol; diurnal; automated sampling

向作者/读者索取更多资源

This study assessed the performance and safety of Fluispotter for the first time in healthy subjects. Fluispotter is a novel body-worn device with automated extraction, collection, and storage of venous dried blood spot samples. The majority of participants were able to successfully collect all the samples, demonstrating that Fluispotter is safe and feasible.
Repeated blood sampling is required in certain clinical and research settings, which is currently performed by drawing blood from venous catheters requiring manual handling of each sample at the time of collection. A novel body-worn device for repeated serial samples, Fluispotter & REG;, with automated extraction, collection, and storage of up to 20 venous dried blood spot samples over the course of 20 h may overcome problems with current methods for serial sampling. The purpose of this study was to assess the performance and safety of Fluispotter for the first time in healthy subjects. Fluispotter consists of a cartridge with tubing, a reservoir for flushing solution, pumps and filterpaper, and a multi-lumen catheter placed in the brachial vein. We recruited healthy subjects for testing in an in-hospital setting. Fluispotter was attached by an anesthesiologist to 22 healthy subjects of which 9/22 (40.9%) participants had all 20 samples taken, which was lower than the goal of complete sampling in 80% of the subjects (P = 0.02). The main reason for sample failure was clogging of blood flow which was observed in 11/22 (50%) of the participants. No serious adverse events occurred, and the participants rated the pain from the insertion and the removal of catheter as very low. A cortisol profile showed nadir values at midnight and highest values at 05:00 h. Although full sampling was not successful in all participants, the Fluispotter technology proved safe and highly acceptable to the participants producing the expected cortisol profile without the requirement of staff during sample collection.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据