4.6 Article

The Cyclic Fatigue Resistance of Different Lengths of CM Gold Wire and CM Blue Wire NiTi Alloy Endodontic Rotary Files: An In Vitro Study

期刊

APPLIED SCIENCES-BASEL
卷 13, 期 7, 页码 -

出版社

MDPI
DOI: 10.3390/app13074612

关键词

CM Blue wire; CM Gold wire; cyclic fatigue; endodontics; length; NiTi alloy

向作者/读者索取更多资源

The objective of this study was to measure and compare the impact of the length of CM Gold Wire and CM Blue Wire NiTi alloy endodontic rotary files on their resistance to cyclic fatigue. The results showed significant differences among different length groups, with file length being inversely proportional to cyclic fatigue resistance.
Background: The objective of the present study was to measure and compare how the length of CM Gold Wire and CM Blue Wire NiTi alloy endodontic rotary files impacts their resistance to cyclic fatigue. Methods: A total of 40 sterile endodontic rotary files were chosen and allocated to the following study groups: (A) 25.06 CM Gold wire NiTi alloy endodontic rotary files, 31 mm in length (n = 10); (B) 25.06 CM Gold wire NiTi alloy endodontic rotary files, 25 mm in length (n = 10); (C) 25.06 CM Gold wire NiTi alloy endodontic rotary files, 21 mm in length (n = 10); and (D) 25.06 CM Blue wire NiTi alloy endodontic rotary files, 17 mm in length (n = 10). A specialized device was designed using artificial root canal systems made from stainless steel for the dynamic cyclic fatigue tests, with an apical diameter of 250 mu m, curvature angle of 60 degrees, radius of curvature of 5 mm, lengths of 31, 25, 21, and 17 mm, and a 6% taper. An individual operator determined failure of the endodontic rotary instrument through direct observation and the tests were filmed so as to precisely measure the exact time to failure. The results were analyzed using ANOVA and Weibull statistical analysis. Results: The results found statistically significant differences across all study groups (p < 0.05). Conclusions: Rotary file length is inversely proportional to the cyclic fatigue resistance of the 25.06 CM Gold wire NiTi alloy at 31 mm, 25 mm, and 21 mm in length and of the 25.06 CM Blue wire NiTi alloy 17 mm length endodontic rotary files, with a greater length contributing to lower resistance to cyclic fatigue.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据