Validation of a rapid SARS-CoV-2 antibody test in general practice

期刊

BMJ OPEN
卷 13, 期 5, 页码 -

出版社

BMJ PUBLISHING GROUP
DOI: 10.1136/bmjopen-2022-069997

关键词

-

向作者/读者索取更多资源

A study validated a rapid serological test for SARS-CoV-2 antibodies in healthcare providers in Belgium. The results showed that the test underestimated or overestimated the true seroprevalence in the population.
Objectives To validate a rapid serological test (RST) for SARS-CoV-2 antibodies used in seroprevalence studies in healthcare providers, including primary healthcare providers (PHCPs) in Belgium. Design A phase III validation study of the RST (OrientGene) within a prospective cohort study. Setting Primary care in Belgium. Participants Any general practitioner (GP) working in primary care in Belgium and any other PHCP from the same GP practice who physically manages patients were eligible in the seroprevalence study. For the validation study, all participants who tested positive (376) on the RST at the first testing timepoint (T1) and a random sample of those who tested negative (790) and unclear (24) were included. Intervention At T2, 4 weeks later, PHCPs performed the RST with fingerprick blood (index test) immediately after providing a serum sample to be analysed for the presence of SARS-CoV-2 immunoglobulin G antibodies using a two-out-of-three assay (reference test). Primary and secondary outcome measures The RST accuracy was estimated using inverse probability weighting to correct for missing reference test data, and considering unclear RST results as negative for the sensitivity and positive for the specificity. Using these conservative estimates, the true seroprevalence was estimated both for T2 and RST-based prevalence values found in a cohort study with PHCPs in Belgium. Results 1073 paired tests (403 positive on the reference test) were included. A sensitivity of 73% (a specificity of 92%) was found considering unclear RST results as negative (positive). For an RST-based prevalence at T1 (13.9), T2 (24.9) and T7 (70.21), the true prevalence was estimated to be 9.1%, 25.9% and 95.7%, respectively. Conclusion The RST sensitivity (73%) and specificity (92%) make an RST-based seroprevalence below (above) 23% overestimate (underestimate) the true seroprevalence.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据