Micro-CT Evaluation of Microgaps at Implant-Abutment Connection

期刊

MATERIALS
卷 16, 期 12, 页码 -

出版社

MDPI
DOI: 10.3390/ma16124491

关键词

dental implant; microgap; micro-computed tomography; abutment

向作者/读者索取更多资源

The aim of this study was to evaluate the size of microgaps between prefabricated and customised abutments (Astra Tech, Dentsply, York, PA, USA; Apollo Implants Components, Pabianice, Poland) mounted on a standard implant. The measurement of the microgap was performed using micro-computed tomography (MCT). The results showed that most samples did not exhibit any microgaps, and the microgap size of the Apollo abutment was higher and more variable than that of the Astra one.
The assessment of microgaps at the implant-abutment interface is an important factor that may influence clinical success. Thus, the aim of this study was to evaluate the size of microgaps between prefabricated and customised abutments (Astra Tech, Dentsply, York, PA, USA; Apollo Implants Components, Pabianice, Poland) mounted on a standard implant. The measurement of the microgap was performed using micro-computed tomography (MCT). Due to 15-degree rotation of samples, 24 microsections were obtained. Scans were performed at four levels established at the interface between the abutment and the implant neck. Moreover, the volume of the microgap was evaluated. The size of the microgap at all measured levels varied from 0.1 to 3.7 & mu;m for Astra and from 0.1 to 4.9 & mu;m for Apollo (p > 0.05). Moreover, 90% of the Astra specimens and 70% of the Apollo specimens did not exhibit any microgaps. The highest mean values of microgap size for both groups were detected at the lowest portion of the abutment (p > 0.05). Additionally, the average microgap volume was greater for Apollo than for Astra (p > 0.05). It can be concluded that most samples did not exhibit any microgaps. Furthermore, the linear and volumetric dimensions of microgaps observed at the interface between Apollo or Astra abutments and Astra implants were comparable. Additionally, all tested components presented microgaps (if any) that were clinically acceptable. However, the microgap size of the Apollo abutment was higher and more variable than that of the Astra one.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据