Comparison of corneal measurements in normal and keratoconus eyes using Anterior Segment Optical Coherence Tomography (AS-OCT) and Pentacam HR topographer

期刊

BMC OPHTHALMOLOGY
卷 23, 期 1, 页码 -

出版社

BMC
DOI: 10.1186/s12886-023-02946-w

关键词

Topographic Corneal Thickness; Keratoconus; Pentacam; AS-OCT

向作者/读者索取更多资源

This study compared the measurements of K, CCT, and TCT obtained by Optovue AS-OCT and Wavelight Oculyzer Pentacam HR in keratoconus (KC) eyes and normal eyes. The results showed significant differences between the two groups in terms of BCVA, intraocular pressure, and CCT. There were also significant differences in TCT measurements between the groups using Pentacam HR and AS-OCT. Both imaging techniques provided accurate identification of KC eyes and healthy ones, but there was a significant difference in K readings.
BackgroundKeratoconus (KC) is easily recognized by its unique topographic pattern, but it can be difficult to distinguish subclinical form of the disease from the normal cornea. Optovue anterior segment optical coherence tomography (AS-OCT) helps diagnose KC.Aim of the workTo assess and the level of agreement of Keratometry-readings (K), Central Corneal Thickness (CCT) and Thinnest Corneal Thickness (TCT) measurements obtained by Optovue AS-OCT and Wavelight Oculyzer Pentacam HR in two groups: KC eyes and normal eyes.Patients and methodsThis is a prospective clinical observational study. The study included 110 eyes divided into two groups. The study group included 62 eyes with topographic evidence of KC. The control group included 48 eyes of normal subjects with no topographic evidence of KC. All of the participants underwent full cycloplegic refraction, spectacle best-corrected distance visual acuity, comprehensive slit-lamp biomicroscopy and fundoscopy. All participants underwent corneal topography by Pentacam HR and AS-OCT.ResultsThere were highly significant differences between the studied groups as regarding BCVA, intraocular pressure and CCT measurements which were found to be lower among KC group compared to the control one. There were highly significant differences between the studied groups regarding TCT measurement detected by Pentacam HR and AS-OCT which was found to be lower among the keratoconus group compared to the control one (470.9, 455.7 versus 541.9 and 518.7 respectively).ConclusionBoth Scheimpflug-based imaging and AS-OCT provide comparable readings with a good agreement regarding corneal pachymetry in keratoconus group with accurate identification of KC eyes and healthy ones. However, there was a significant difference in K readings between both devices in Keratoconus and control group.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据