4.4 Article

Oscillometric pulse wave analysis for detecting low flow arteriovenous fistula

期刊

BMC NEPHROLOGY
卷 24, 期 1, 页码 -

出版社

BMC
DOI: 10.1186/s12882-023-03243-w

关键词

Pulse wave analysis; Haemodialysis; arteriovenous fistula; Duplex sonography

向作者/读者索取更多资源

Convenient oscillometric devices can be used to detect blood flow below 500 ml/min in a real-life clinical setting. Among the parameters measured by these devices, the relative volumetric change (AMP) showed the highest diagnostic accuracy.
Background Pulse wave analysis may be useful to assess fistula function. We aimed to prospectively evaluate if convenient oscillometric devices are applicable to detect flow below 500 ml/min in a real life clinical setting. Methods Pulse waves were recorded ambilaterally with the vicorder((R)) device at the brachial artery in 53 patients on haemodialysis with native fistula. Primary variables consisted of the mean slope between the systolic maximum and the diacrotic notch (Slope2), the sum of the mean slopes in the four characteristic sections of pulse waves (Slope Sigma) and the amplitude of relative volumetric change in the measuring cuff at the upper arm (AMP). Fistula flow was measured with the use of duplex sonography using a standardized approach. Results Parameter values above or below the median indicated measurement at the non-fistula side, with sensitivities/specificities of 0.79/0.79 (p < 0.001) for Slope 2, 0.64/0.64 (p = 0.003) for Slope Sigma and 0.81/0.81 (p < 0.001) for AMP if measurements at the fistula and non-fistula arm were considered. ROC-analyses of parameter values measured at the fistula to detect low flow demonstrated AUCs (with CI) of 0.652 (0.437-0.866, p = 0.167) for Slope2, 0.732 (0.566-0.899, p = 0.006) for Slope Sigma and 0.775 (0.56-0.991, p = 0.012) for AMP. The point with maximal youden's index was regarded as optimal cut-off, which corresponded to sensitivities and specificities of 0.8/0.56 for slope2, 0.86/ 0.56 for Slope Sigma and 0.93/0.78 for AMP. Conclusion Functional surveillance with oscillometry is a promising clinical application to detect a low fistula flow. Among all investigated pulse wave parameters AMP revealed the highest diagnostic accuracy.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据